CS320A Technical specifications

 
Analog inputs and acquisition parameters
Channels:2
Sample rate:100MS/sec. for each channel
Resolution:10 bits (12 bits or 14 bits optional)
Sample memory:4M samples(8M samples optional)
Input impedance:1MOhm ±2% / 20pF ±3pF
Maximum voltage:300Vrms, 420V peak (no damage)
(duty cycle <50%, pulse width <100 msec)
Input coupling:DC, AC, GND
Acquisition modes:Single Shot, Triggered, Automatic, Repetitive (High Frequency), Multiple Frame
Acquisition rate:Acquisition rate to PC via USB is 10 Frames/sec.
Time delay:
(between channels)
200ps (typical)
Voltage and time base
Full Scale Volts Range:from 20mV to 20V (1x probe)
Voltage base sensitivity:0.02mV for 20 mV full scale
Maximal analog bandwitdh:100 MHz, -3dB
Lower Frequency limit, AC coupled:10Hz (probe 1x), 1Hz (probe 10x)
Rise time at the BNC:<3ns (typical)
Peak detect response:captures all pulses >10ns in duration
Sample Rate Range:from 1500 samples/sec. to 100M samples/sec
Time base range:from 1ns/div to 5s/div in 1,2,5 sequence
Sample Clock jitter:1 ps rms (typical)
Record Length:1024 – 2 000 000 samples for each channel
Waveform interpolation:Sin(x)/x
Captured Sample window duration:1usec – 40msec with 10ns resolution
40msec – 42.9 secs with 10ns - 10usec resolution.
(Lower sample rates are used for smaller capture buffer sizes)
Trigger possibilities
Trigger Sensitivity, Edge Triggered:Analog Channels: 0.2 Div from DC to 50 MHz
Analog Hardware Trigger: 0.5 div from DC to 200 MHz
External Trigger: 50 mV from DC to 100 MHz
Trigger modes:Edge, Window, Pulse Duration
Trigger filtering:Noise reject, HF reject, LF reject
Trigger level range:Internal: defined by scope graph
External: ±20V in 40 mV increments
Trigger holdoff range:0 to 42.9 secs with 10ns resolution
Signal Generator Plug-in CS700 (optional extension)
Generated functions:Sine, Square or Triangle output signals, possibility of sweep and modulation funtions
Generation Method:Direct Digital Synthesis
Output Sampling Rate:50MS/sec.
Frequency Range:Sine, Square: 0.2 Hz to 10 MHz; Triangle: 0.2 to 1 MHz
Output Voltage Range:from 100mV to 5VPEAK-PEAK including DC offset with 10mV resolution
Output Voltage Offset Range:-4V to +4V with 10mV resolution
Output Impedance:50 Ohm
Signal to Noise Ratio:-60 dBc (typical)
Total Harmonic Distortion:for frequencies from 0 to 1 MHz: < -65 dBc: ; for frequencies >1MHz: < -35 dBc
General specifications
Probe Compensator Output:3V, 1kHz into >100k Ohm load
Power supply:6-20V DC
Power Consumption:5W
Temperature range:Operating: 0°C to +40°C
Storage: -20°C to +60°C
Dimensions:35mmx153mmx195mm (HeightxWidthxDepth)
Minimal PC requirements:PC with Pentium II processor
memory 64MB at least
free USB port
Package content:CS320A device
2 oscilloscope probes 250MHz (CS1001)
Power supply adaptor 230V/50Hz
Interconnecting USB cable
Printed documention
CD with software, documentation, etc.
 

Complete detail technical specification for Cleverscope CS320A
can be downloaded in PDF format HERE

 
.